Technical parameters of the whole machine
Equipment capacity: 2kVA
Input voltage: 220V
Rated input current: 9.1A
High voltage output: 0-20kV
Rated output current: 2.000mA
Resolution: 0.1uA
Maximum electrode opening distance: 50mm
Maximum electrode diameter: ¢150mm
Equipment inherent leakage: 0.12mA/20kV (basically linear)
Timing range: 0-999S
Ambient temperature: -20°C to 50°C
Voltage accuracy: ≤1.5% (F.S) ± 5 words
Current accuracy: ≤1.5% (F.S) ± 5 words
Features
1. 320×240 liquid crystal display, high-speed thermal printer.
2. High-voltage voltage, high-voltage leakage current, low-voltage voltage, low-voltage current 4-channel measurement methods, high-precision sensors and high-performance 14-bit AD acquisition chip.
3. The full keyboard operation mode of man-machine dialogue, the whole process of intelligent work, you can choose between automatic mode and manual mode.
4. Real-time display of high voltage voltage, high voltage leakage current, low voltage voltage, low voltage current, time and withstand voltage results, the display is intuitive and clear.
5. Complete over-voltage and over-current protection, output voltage, high-voltage current upper limit, low-voltage current upper limit and timing time can be set arbitrarily.
6. The zero return detection function, the test can be carried out after the zero return is confirmed, which is safe and reliable.
7. With insulation resistance test function, it directly reflects the insulation strength of the tested product.
8. Approximate voltage regulation algorithm, automatically withstand voltage timer after reaching the set voltage, and automatically drop the voltage back to zero after the timer ends.
9. When the set high-voltage current or low-voltage current is exceeded, the voltage output will be automatically cut off, the voltage will be reduced to zero, and an audible and visual alarm will occur.
10. Sophisticated software and hardware anti-jamming design, multiple anti-jamming methods, adapt to harsh electromagnetic environment.