Technical index
1. Working voltage: AC220V, 50Hz
2. Instrument power: 50W
3. Instrument accuracy: 0.2 level
4. Pressure display resolution: 0.001MPa
5. Pressure calibration range: 0~1MPa
6. Ambient temperature display resolution: 0.1°C
7. Measuring range of ambient temperature: -20℃~80℃
8. Working humidity: £90%RH
9. Communication method: USB high-speed communication
10. Check object: single alarm, single lockout, single alarm and single lockout, single alarm and double lockout
11. Display mode: 320×240 graphic dot matrix monochrome LCD screen
12. Operation mode: Rotate the mouse (one-key shuttle) to input
13. Printer type: high-speed micro printer
14. Printing method: Chinese report printing
15. Storage capacity: 50 sets of test results can be stored at the same time
16. Dimensions: Host 365×300×220(mm3)
Accessory box 400×370×200(mm3)
17. SF6 gas density display mode: the pressure under the measured environment, the equivalent pressure at 20°C.
18. Instrument weight: 8kg
Performance characteristics
1. This product uses a 32-bit microprocessor and TI's high-speed signal processing chip for detection and control, with a high degree of integration. Mechatronics design, high precision, good repeatability and high reliability.
2. Equipped with a large Chinese LCD screen and a rotating mouse (one-button shuttle) man-machine interface, simple operation, beautiful interface, and all parameters and status are clear at a glance.
3. Automatically complete pressure measurement and 20°C value conversion, thereby completing dynamic automatic compensation between pressure and temperature. It also displays the pressure at the measured ambient temperature, the pressure at the ambient temperature at 20°C, and the ambient temperature. Thoroughly solve the problem of difficult field calibration of SF6 gas density relay.
4. All testing processes are automatically completed by the instrument without manual intervention, which avoids tedious manual airway adjustment operations.
5. All-metal quick-plug connectors are used to effectively avoid the locking phenomenon of plastic connectors in low temperature environments.
6. The test results are printed as a report of Chinese characters, and the test results are intelligently analyzed.
7. It can store 50 sets of test results at the same time, and has a power-down data protection function, and the previous test results can be inquired and printed at any time.
8. There is no need for a constant temperature room during the calibration process, and the SF6 gas density relay and pressure gauge can be calibrated within any effective temperature range.
9. Equipped with various types of transition joints, the density relays of most types of switches can be verified on site without disassembly.
During the calibration process, the gas consumption is very small, SF6 gas is not wasted, the test cost is low, and it does not pollute the environment.
10. With the function of online modification of the system clock.
11. It has the function of online modification of system pressure, which can be modified on site.
12. The core components are imported components with reliable performance.
13. It can test one set of alarm signals and two sets of blocking signals at the same time.
14. This product is a portable tool, easy to use and reliable, it is the best choice for SF6 density relay calibration.