Main Specifications
AC output
·Voltage range 200mV, 400mV, 2000m V, 4000m V, 20V, 40V, 100V, 200V, 400V, 1000V;
·Current range 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A;
·Adjustment range: 0~120% (except for 1000V voltage);
·Adjust the fineness to 5×10-5;
·Voltage and current output capacity up to 20VA;
·Voltage and current setting accuracy
A: Voltage 20~1000V, current 0.2~20A range ±(0.03%RD+0.02%RG);
B: Voltage 200~4000mV, current 2~40mA range ±(0.3%RD+0.2%RG);
·Voltage and current output stability A: ≤0.01% /60s (except for gears below 40mA and 4000mV);
·Voltage and current waveform distortion degree ≤0.3% (except for gears below 40mA and 4000mV);
·Sine wave output frequency adjustment range (45~65) ·N Hz (N=1,2,3,……31);
·Sine wave output frequency adjustment fineness 0.001·N Hz (N=1, 2, 3, ... 31);
·The accuracy of sine wave output frequency setting value is 0.01 Hz;
·Phase adjustment range 0~359.99°;
Adjust the fineness to 0.01°;
·Harmonic output 2~31 times, amplitude 0~20% (rated value of fundamental wave);
Phase fineness 0.01°·N (N is the harmonic order, N= 2, 3, 4,…… 31);
The output can contain multiple harmonics at the same time, or you can set the fundamental wave amplitude to 0 to output a sine wave higher than the power frequency.
DC output
·Voltage range 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V;
·Current range 20μA, 40μA, 200μA, 400μA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A,
5A, 10A, 20A;
·Voltage and current adjustment range 0~110% (except for 1000V voltage);
·Voltage and current adjustment fineness 5×10-5;
·Voltage and current output capacity up to 20W;
·Voltage adjustment accuracy ±(0.03%RD+0.02%RG);
·Current setting accuracy ±(0.06%RD+0.04%RG);
·Voltage and current output stability ≤0.01%/60 (peak-peak value) s;
·Voltage and current output ripple ≤1%.
Point frequency square wave
·Point frequency square wave output 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz;
· The amplitude of the square wave is 4.5V.
·Setting accuracy 5×10-5;
Electronic resistance box output
·Range 0~200Ω, 0~2kΩ, 0~20kΩ, 0~200kΩ, 0~2MΩ;
·Adjustment fineness 0.025%;
·The setting accuracy is ±(0.3%RD + 0.2%RG) at 0~200Ω, 0~2MΩ gear;
Except 0~200Ω, 0~2MΩ range ±(0.12%RD + 0.08%RG).
Working power: Single-phase 220V±10%, 50Hz±5%.