Main Specifications
AC output
·Voltage range 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V;
·Current range 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A;
·Adjustment range 0~120% (except for 1000V voltage);
·Adjust the fineness to 5×10-5;
·Voltage and current output capacity up to 20VA;
·Voltage and current setting accuracy
A: Voltage 20~1000V, current 0.2~20A range ±(0.03%RD+0.02%RG);
B: Voltage 200~4000mV, current 2~40mA range ±(0.3%RD+0.2%RG);
·Voltage and current output stability A: ≤0.01% /60s; B: ≤0.05% /60s;
·Voltage and current waveform distortion degree ≤0.3%;
·Sine wave output frequency adjustment range (45~65)·N Hz (N=1,2,3,……31);
·Sine wave output frequency adjustment fineness 0.001·N Hz (N=1, 2, 3, ... 31);
·Sine wave output frequency adjustment accuracy 0.01Hz;
· Harmonic output 2~31 times, amplitude 0~20% (rated value of fundamental wave);
Phase fineness 0.01°·N (N is the harmonic order, N= 2, 3, 4,…… 31);
The output can contain multiple harmonics at the same time, or you can set the fundamental wave amplitude to 0 to output a sine wave higher than the power frequency.
DC output
·Voltage range 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V;
·Current range 20μA, 40μA, 200μA, 400μA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A,
5A, 10A, 20A;
·Voltage and current adjustment range 0~110% (except for 1000V voltage);
·Voltage and current adjustment fineness 5×10-5;
·Voltage and current output capacity up to 20W;
·Voltage adjustment accuracy ±(0.03%RD+0.02%RG);
·Current setting accuracy ±(0.06%RD+0.04%RG);
·Voltage and current output stability ≤0.01%/60s;
·Voltage and current output ripple ≤1%.
Frequency output
·Sine wave output range (provided by AC voltage and current output) (45~65) ·N Hz;
·Fineness 0.001·N Hz (N=1, 2, 3, ... 31);
·The setting accuracy is 5×10-5.
Point frequency square wave
·Point frequency square wave output 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz;
· The amplitude of the square wave is 4.5V.
·Setting accuracy 5×10-5;
Electronic resistance box output
·Range 0~200Ω, 0~2kΩ, 0~20kΩ, 0~200kΩ, 0~2MΩ;
·Adjustment fineness 0.025%;
·The setting accuracy is ±(0.3%RD + 0.2%RG) at 0~200Ω, 0~2MΩ gear;
Except 0~200Ω, 0~2MΩ range ±(0.12%RD + 0.08%RG).
Working power: Single-phase 220V±10%, 50Hz±5%.
Use environment temperature 20℃±10℃, relative humidity ≤85%RH.
Volume and weight 440×460×165mm3, about 12kg.
Main functions and features
1. The device is composed of a single-phase AC voltage standard source, a current standard source, a DC voltage standard source, and a current standard source.
2. With calibration function, it can correct the output value of the standard source according to the displayed value of the external standard.
3. The clamp-on ammeter can be tested through the specially designed terminal button and the specially equipped "clamp-on-meter check coil", up to 1000A.
4. It can manually and semi-automatically test the digital multimeters, pointer multimeters, clamp-on ammeters, and installation digital meters of class 0.2 and below.
Includes digital meters with DC analog output.
5. It can generate distortion waves containing 2 to 31 harmonics. The number, order, amplitude and phase of the harmonics to the fundamental wave can be controlled by program.
6. Equipped with RS-232 interface or USB interface and large-capacity non-volatile memory, which can store the test data of 500 pieces of tested meters.
Through the upper computer software, the device can be controlled by the PC to perform automatic and manual inspections, and process and manage the results.